Figure
3: PcCr 1417 Dimensions
Electrical Requirements
50 cm
80 cm
100 cm
Voltage:
90 V–250 V AC
Frequency:
50Hz–60Hz
Fuses:
Two fuses are located next to the On/Off switch at the
back of the scanner. Each fuse protects one of the
power supplies. The fuses are 1A (slow blow).
UPS:
Online; 500 VA
3.5 Environmental Requirements
3.5.1 Temperature
Maintain the room temperature between 18ºC and 30ºC, normal
lab conditions; relative humidity 80%, non-condensing.
3.5.2 Light
The PcCR 1417 is sealed and protected from light penetration.
However, it is recommended to keep the PcCR 1417 in standard
room light, away from exposure to direct sunlight.
The PcCR 1417 uses phosphor plates for acquiring images. The
phosphor plates are sensitive to light, and therefore exposure to
light should be kept to a minimum. The phosphor plates should
Site Preparation 3-2