FieldFox
Warranted performance. Specifications include guardbands to
account for the expected statistical performance distribution,
measurement uncertainties, and changes in performance due to
environmental conditions. The following conditions must be met:
FieldFox has been turned on at least 10 minutes (unless other-
wise stated)
stated)
temperature range, unless otherwise indicated; does not include
guardbands. It is not covered by the product warranty. The
FieldFox must be within its calibration cycle.
A general, descriptive term or design parameter. It is not tested,
and not covered by the product warranty.
Specifications
Frequency
Frequency range
2 MHz
1
to 4 GHz
2 MHz
2
to 6 GHz
Frequency reference
Accuracy
Aging rate
Temperature stability
Frequency resolution
2 MHz to 1.6 GHz 2.5 kHz
> 1.6 GHz to 3.2 GHz 5 kHz
> 3.2 GHz to 6 GHz 10 kHz
Measurement speed
Return loss 1.5 ms/point (nominal) 1.75 GHz to 3.85 GHz, 1001 points, Cal ON
2.4 ms/point (nominal) 0 to 500 ft, 601 points, Cal ON
101, 201, 401, 601, 801, 1001 (up to 10,001 using SCPI)
> 42 dB
QuickCal > 42 dB (typical)
3
Source match
> 36 dB
QuickCal ≥ 35 dB (typical)
3
Reflection tracking
0.06 dB
QuickCal
0.15 dB (typical)
3
A condensed version of the specifications is provided here. See the User’s Guide for the complete version;
http://cp.literature.agilent.com/litweb/pdf/N9912-90001.pdf
1. Spectrum analyzer (Option 230) start frequency is 100 kHz, usable to 5 kHz.
2. Spectrum analyzer (Option 231) start frequency is 100 kHz, usable to 5 kHz.
3. Requires 90 minute warm up